• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 BROVIAC CV CATHETER REPAIR KIT, SINGLE-LUMEN, WHITE, 4.2F; PORT AND CATHETER ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 BROVIAC CV CATHETER REPAIR KIT, SINGLE-LUMEN, WHITE, 4.2F; PORT AND CATHETER ACCESSORIES Back to Search Results
Model Number 0601610
Medical Device Problem Codes Fluid/Blood Leak (1250); Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/07/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a catheter placement procedure, the catheter allegedly fractured.There was no reported patient injury.
 
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.(expiration date: 08/2022).
 
Additional Manufacturer Narrative
Manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.A definitive root cause for the reported failure could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.D4 (expiration date: 08/2022).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during a catheter placement procedure, the catheter allegedly fractured.It was further reported that the fluid allegedly leaked.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BROVIAC CV CATHETER REPAIR KIT, SINGLE-LUMEN, WHITE, 4.2F
Common Device Name
PORT AND CATHETER ACCESSORIES
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key14193974
Report Number3006260740-2022-01476
Device Sequence Number11845280
Product Code LJS
UDI-Device Identifier00801741036385
UDI-Public(01)00801741036385
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K830406
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0601610
Device Catalogue Number0601610
Device Lot NumberREEY3113
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/30/2022
Supplement Date Received by Manufacturer05/31/2022
Initial Report FDA Received Date04/25/2022
Supplement Report FDA Received Date06/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/07/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-