| Model Number |
100/870/080CZ |
| Medical Device Problem Code |
Unintended Deflation (4061)
|
| Health Effect - Clinical Code |
Ventilator Dependent (2395)
|
| Date of Event |
03/25/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
One used decontaminated sample 007/910/880 8mm blu suctionaid sub-assy was returned for investigation without its original packaging.Under visual inspection the sample appeared to be in good condition.During manufacturing process, the devices are 100% inflation tested, which includes inflating each device cuff and leaving for a 12 hour period.Reductions in pressure over this time are considered a failure and the device would be rejected.The inflation test was repeated on the received sample.It was confirmed that after 12 hours the cuff was still fully inflated.Based on results of testing the reported failure was not observed.No trend of similar customer complaints was identified.A dhr was not completed since no fault was found.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient had a leaking tracheostomy tube cuff.The patient required an earlier tracheostomy change.Upon decannulation, the hcp noted that a 'build of bubble' was noted within the cuff.No injury to the patient.
|
| |
|