| Model Number |
TGM454515 |
| Medical Device Problem Code |
Material Perforation (2205)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/29/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Images were returned for evaluation.The evaluation stated: one time-point available for evaluation: post-implantation cta dated (b)(6) 2022.Comparison series axial images appear to show a possible lack of proximal device apposition on the inner curve.Comparison axial images does not show an increase in pooling contrast on the delayed contrast series, thereby, ruling out a type ii endoleak.There appears to be bright contrast outside the implanted device communicating with contrast inside the device ~24mm, by outer curve length, distal to the proximal circumferential end of the device.Cannot confirm communication between the small amount of contrast outside the proximal device and the contrast pooling distally.(~24mm distal).Probable type iii endoleak.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2021, the patient was treated for kommerell divirticulum/ aberrant sublclavian.The physician implanted a gore® tag® conformable thoracic stent graft with active control system.The patient tolerated the procedure.On (b)(6) 2022, the gore fsa received a scan from the physician that appears to be a possible type iii endoleak.Images were received for evaluation.As of this time, the physician has not taken any action to address the endoleak.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was received that a reintervention was performed to repair the reported endoleak.No endoleak was seen on angiogram images.Event type was changed from product malfunction to serious injury.
|
| |
|
Search Alerts/Recalls
|