As reported, an advance 18 lp low profile balloon catheter ruptured during a procedure involving a lesion of the left superficial femoral artery.The vessel was calcified.The balloon ruptured following a second inflation to nominal pressure using an inflation device.No blood was noted in the inflation device following the rupture.The device was removed from the patient through a sheath, and a new balloon was used to continue to procedure.A section of the device did not remain inside the patient¿s body.The patient did not require any additional intervention due to this occurrence.The patient did not experience any adverse effects due to this occurrence.
|
|
Summary of event: as reported, an advance 18 lp low profile balloon catheter ruptured during a procedure involving a lesion of the left superficial femoral artery.The vessel was calcified.The balloon ruptured following a second inflation to nominal pressure using an inflation device.No blood was noted in the inflation device following the rupture.The device was removed from the patient through a sheath, and a new balloon was used to continue to procedure.A section of the device did not remain inside the patient¿s body.The patient did not require any additional intervention due to this occurrence.The patient did not experience any adverse effects due to this occurrence.Investigation evaluation: reviews of the complaint history, device history record, drawing, instructions for use (ifu), specifications, and quality control procedures were conducted during the investigation.The complaint device was not returned to cook for investigation.A document-based investigation evaluation was performed.No related non-conformances were found, and there have been no other reported complaints for this lot number.The product ifu warns ¿apply negative pressure to lumen labeled ¿balloon¿ prior to introduction.Advance the balloon dilation catheter counter-clockwise over a per-positioned.018 (0.46 mm) wire guide.¿ the information provided upon review of the dmr, dhr, and ifu suggests that there is evidence the device was manufactured to specification.There is no evidence of non-conforming devices in-house or in the field.Cook has concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.Based on the information provided and the results of the investigation, cook has concluded that the patient¿s calcified anatomy contributed to this event.The risk analysis for this failure mode was reviewed and no additional escalation was required.The appropriate personnel have been notified and cook will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|