MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS IOL; INTRAOCULAR LENS Back to Search Results
Model Number DFR00V
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/07/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during implantation the trailing haptic of the pre-loaded intraocular lens (iol) sheared off.Through follow-up additional information was received stating the pre-loaded iol was fully inserted into the patient's ocular dexter (right eye).There was no incision enlargement, no sutures, and no vitrectomy.The same procedure was completed successfully using a backup lens with the same model and diopter size.No further information is available.
 
Additional Manufacturer Narrative
Additional information: patient weight: unknown as information was asked but not provided.Date implanted: not applicable, as the iol was removed and replaced during the same procedure.Date explanted: not applicable, as the iol was removed and replaced during the same procedure.The device was not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Product investigation was completed, and the following fields were updated accordingly: section d-9: device available for evaluation: yes.Section d-9: date returned to manufacturer: 02-may-2022.Section h-3: device evaluated by manufacturer: yes.Device evaluation: the complaint lens was received inside of a specimen cup.Visual inspection under magnification revealed that the lens was received cut in half and with a haptic detached.The lens was cleaned and, scratches were observed on the lens.Inspection of the handpiece revealed that the missing haptic was stuck inside of the cartridge and that it was overridden by the plunger rod, which is consistent with a haptic that was torn off during implantation.The cartridge tip was also observed to be damaged.The handpiece was disassembled, and the assembly was inspected; no assembly issues were observed.Complaint issue "haptic detached" was identified during product evaluation; however, based on the complaint investigation results the complaint issue could not be confirmed to be related to the manufacturing or design process.Therefore, there is no indication of a product deficiency or product malfunction.Conclusion: based on the investigation results, there is no indication of a product malfunction or product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Corrected data: section h6: as part of an internal review of our mdrs it was identified that health effect - impact code 2199 provided on the initial report needs to be corrected.The correct impact code is health effect - impact code: 4631 iol removal and replacement.The following section has been updated accordingly: section h-6 health effect - impact code: 4631 iol removal and replacement.Removed health effect - impact code 2199 no health consequences or impact code as it is not applicable.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key14196441
Report Number3012236936-2022-01096
Device Sequence Number4446977
Product Code HQL
UDI-Device Identifier05050474750746
UDI-Public(01)05050474750746(17)240524
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/24/2024
Device Model NumberDFR00V
Device Catalogue NumberDFR00VU235
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/07/2022
Supplement Date Received by Manufacturer07/14/2022
08/26/2022
Initial Report FDA Received Date04/25/2022
Supplement Report FDA Received Date07/25/2022
09/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/24/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age63 YR
Patient SexMale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-