• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DRAEGER MEDICAL SYSTEMS, INC. INCUBATOR C2000; INCUBATOR, INFANT, STATIONARY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DRAEGER MEDICAL SYSTEMS, INC. INCUBATOR C2000; INCUBATOR, INFANT, STATIONARY Back to Search Results
Model Number C2000
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Fall (1848); Hematoma (1884)
Date of Event 04/11/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a registered nurse (rn) at the customer site was prompted to the patient's room after hearing the patent's mother in distress.Upon entering the room, the rn found the patient's mother picking up the patient from the floor directly in front of the incubator.The patient's mother alleged that the incubator back door was not locked and that the incubator door fell open and the patient fell from the incubator onto the floor.There were no witnesses in the room with the mother when the incident occurred.The patient was diagnosed with a hematoma to the top right back of the head.Treatment from the emergency room physician and rn was administered, and it was decided to medevac the patient to a higher level of care for further assessment.The device was confirmed to be in working order as designed at the time of the event.
 
Additional Manufacturer Narrative
A follow-up report will be submitted upon completion of this investigation.
 
Additional Manufacturer Narrative
The patient's weight was provided (3.48kg); however the customer declined to provide other patient information (i.E.Initials, gender, status).At the time of the event, the device was sequestered and professional engineers from northwest territories health and social services authority were contacted to investigate the device/incident.During validation testing it was confirmed that no malfunction was identified and that the device was safe to return to service.The incubator c2000 instructions for use (ifu) provides the following in the "hood/shell functional checkout" procedure under "check the access panels" - "close the access panel, and rotate both latches until fully engaged; at the controller screen, check that the access panel open symbol turns off." caution - risk of injury - both locking knobs must be fully engaged to avoid accidental opening of the panels; warning - risk of death or serious injury - positively secure all hand port latches and access panel locking knobs to avoid accidental opening; warning - risk of death or serious injury - for infant safety, do not leave the infant unattended when the access panels or hand ports are open.Of note, the access (port) doors are designed to withstand a horizontal force of 2.0 - 2.5kg (20 - 25n).H3 other text : see h10.
 
Event or Problem Description
It was reported that a registered nurse (rn) at the customer site was prompted to the patient's room after hearing the patent's mother in distress.Upon entering the room, the rn found the patient's mother picking up the patient from the floor directly in front of the incubator.The patient's mother alleged that the incubator back door was not locked and that the incubator door fell open and the patient fell from the incubator onto the floor.There were no witnesses in the room with the mother when the incident occurred.The patient was diagnosed with a hematoma to the top right back of the head.Treatment from the emergency room physician and rn was administered, and it was decided to medevac the patient to a higher level of care for further assessment.The device was confirmed to be in working order as designed at the time of the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INCUBATOR C2000
Common Device Name
INCUBATOR, INFANT, STATIONARY
Manufacturer (Section D)
DRAEGER MEDICAL SYSTEMS, INC.
3135 quarry road
telford PA
Manufacturer (Section G)
DRAEGER MEDICAL SYSTEMS, INC
6 tech drive
andover MA
MDR Report Key14196884
Report Number2510954-2022-00002
Device Sequence Number5479756
Product Code FMZ
UDI-Device Identifier04049098000222
UDI-Public04049098000222
Combination Product (Y/N)N
PMA/510(K) Number
K960980
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Risk Manager
Type of Report Initial,Followup
Report Date (Section B) 06/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberC2000
Device Catalogue NumberMU20500
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/06/2022
Supplement Date Received by Manufacturer04/06/2022
Initial Report FDA Received Date04/25/2022
Supplement Report FDA Received Date06/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/12/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
Patient Weight3 KG
-
-