| Model Number |
7317 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/05/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An investigation is currently underway.Upon completion, the results will be forwarded.
|
| |
|
Event or Problem Description
|
|
The customer reported that they received a call from the operating room and the gauze is shredding when being used on the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record (dhr) was reviewed.For sterilized products, the dhr and the sterilization documents undergo further review prior to release to the distribution center.Prior to a lot¿s release, the lots must be deemed acceptable by passing inspections that are based on a valid sampling plan.Inspectors routinely examine a statistical sample both physically and visually.The reported lot met all the defined acceptance requirements and was released.Inspection records show no failures.On evaluation of the sample, minor fraying could be seen.The reported condition was confirmed.The root cause was determined to be the set screws for the heat roller on the machine were loose.Corrective action performed was the manufacturing associates involved with the process to produce this product have been made aware of the issue and are to be alert to this kind of issue.Samples were sent to the production associates for further awareness and the quality impact to our customers.The set screws for the heat roller on the machine are being monitored for looseness.This complaint will be recorded for tracking and trending purposes.
|
| |
|