• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICROPORT ORTHOPEDICS INC. PROFEMUR® Z STEM PLASMA SPRAYED; HIP COMPONENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROPORT ORTHOPEDICS INC. PROFEMUR® Z STEM PLASMA SPRAYED; HIP COMPONENT Back to Search Results
Model Number PHA00266
Medical Device Problem Code Corroded (1131)
Health Effect - Clinical Codes Pain (1994); Loss of Range of Motion (2032)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This event will be updated once the investigation is complete.Trends will be evaluated.
 
Event or Problem Description
Allegedly, plaintiff profemur total hip system implanted on (b)(6) 2016 in her right hip, on (b)(6) 2021 plaintiff presented pain and lack of mobility, plaintiff needed a revision surgery because the device failed due to corrosion at the neck-stem junction of the device.At revision surgery, it was discovered that the device failed due to corrosion of the oblong taper of the profemur cocr modular neck where it seated in the pocket of the profemur.According to the information received in the report, we do not know if the screws, cup, liner and head were revised.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROFEMUR® Z STEM PLASMA SPRAYED
Common Device Name
HIP COMPONENT
Manufacturer (Section D)
MICROPORT ORTHOPEDICS INC.
5677 airline rd.
arlington TN 38002
Manufacturer (Section G)
MICROPORT ORTHOPEDICS INC.
5677 airline rd.
arlington TN 38002
Manufacturer Contact
5677 airline road
arlington, TN 38002
9018674771
MDR Report Key14197220
Report Number3010536692-2022-00137
Device Sequence Number11882507
Product Code LPH
UDI-Device IdentifierM684PHA002661
UDI-PublicM684PHA002661
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Model NumberPHA00266
Device Catalogue NumberPHA00266
Device Lot Number1654607
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date04/08/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/08/2022
Supplement Date Received by Manufacturer04/08/2022
Initial Report FDA Received Date04/25/2022
Supplement Report FDA Received Date06/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-