SMITH & NEPHEW, INC. SCREW BIOSURE REGENESORB 8MM X 30MM; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE Back to Search Results
Model Number 72204401
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/03/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that during an arthroscopy procedure, the screw biosure regenesorb broke off during the deployment.The broken pieces were removed by arthroscopy instrument.The procedure was completed with non-significant surgical delay using a back-up device.No further complications were reported.
 
Additional Manufacturer Narrative
Internal complaint reference (b)(4).H6: health codes updated.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).H3, h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.There was a relationship found between the device and the reported event.A complaint history review found similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the polymer found that the storage requirements for the material are specified, and the material must comply with provided percentage composition specifications as measured by ash test.One undated photo of the device was provided for review and confirms the reported breakage.Per case details, the broken pieces were retrieved from the patient.The procedure was completed using a back-up device.No patient injuries or adverse consequences were reported.Since no patient injuries are being reported no further clinical/medical assessment is warranted at this time.The complaint was confirmed, and the root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include excessive force on the device, excessive torque on the device, attempted correction of a damaged device, off-axis insertion, improper preparation of the insertion site, or an inadvertent impact event inconsistent with normal use.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.A1: patient identifier is unknown and not "none".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCREW BIOSURE REGENESORB 8MM X 30MM
Common Device Name
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14197985
Report Number1219602-2022-00594
Device Sequence Number4193628
Product Code MAI
UDI-Device Identifier00885554036770
UDI-Public00885554036770
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K142948
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Consumer,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/29/2024
Device Model Number72204401
Device Catalogue Number72204401
Device Lot Number50937630
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/04/2022
Supplement Date Received by Manufacturer08/30/2022
09/30/2022
Initial Report FDA Received Date04/25/2022
Supplement Report FDA Received Date08/31/2022
09/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/29/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-