• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) STYLE 115 SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) STYLE 115 SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number 115-222
Medical Device Problem Codes Material Rupture (1546); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Failure of Implant (1924); Seroma (2069)
Date of Event 03/31/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
 
Event or Problem Description
Patient reported right side rupture.Device remains implanted.
 
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformance's noted.
 
Event or Problem Description
Healthcare professional additionally reported via mri ¿can found small amount of seroma around device.¿ the device has been explanted.
 
Additional Manufacturer Narrative
Device evaluation: visual analysis of the returned device identified: underweight, opening, crease fold, wear abrasion and red particles in the surface of the shell.A microscopic analysis was performed which identify a sharp edge opening assessed as unidentified tear opening.A dimension measurement in the shell was performed which identify the thickness within specification.Based on the device analysis the final assessment is: a opening sharp in the posterior side assessed as unidentified (tear) opening.
 
Event or Problem Description
Patient reported right side rupture.Healthcare professional additionally reported via mri ¿can found small amount of seroma around device.¿ the device has been explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 115 SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
terry ingram
12331-a riata trace parkway
building 3
austin, TX 78727
8479366324
MDR Report Key14198034
Report Number9617229-2022-07041
Device Sequence Number4358474
Product Code FTR
UDI-Device Identifier10888628002340
UDI-Public10888628002340
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Foreign,Consumer,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/30/2022
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/26/2021
Device Model Number115-222
Device Catalogue Number115-222
Device Lot Number2966812
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/01/2022
Supplement Date Received by Manufacturer05/31/2022
08/05/2022
Initial Report FDA Received Date04/25/2022
Supplement Report FDA Received Date06/23/2022
08/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/26/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age59 YR
Patient SexFemale
Patient Weight53 KG
-
-