Catalog Number EUP2015X |
Device Problem
Leak/Splash (1354)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/06/2022 |
Event Type
malfunction
|
Event Description
|
During a procedure an attempt was made to use two euphora rx ptca balloon catheters to treat a simple lesion with no calcification. the devices were being used to pre-dilate the lesion.The devices were inspected with no issues noted. the devices did not pass through a previously deployed stent.Resistance was not encountered when advancing the devices.Excessive force was not used during delivery.It was reported that a balloon leak occurred during balloon inflation.The inflation pressure was 12 atm prior to the leak occurring.The leak occurred on the first inflation.The same issue occurred with both devices.No medical or surgical interventionwas required to prevent a permanent impairment of a function.The event did not lead to or extend patient hospitalization.The patient was reported to be alive with no injury.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Additional information: two mp4 short videos were provided.The first video/ image appears to show the delivery of what is most likely one of the euphora rx balloon to the target lesion.The second short video appears to show that the balloon was positioned more distal in the vessel and on inflation contrast is seen leaking from distal end of balloon ¿ confirming a leak/burst.No images were provided to confirm the vessel morphology, or the full delivery and positioning of the balloon prior to initial inflation.It is not possible to determine the cause of the burst/leak from the images provided.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Product analysis: the device was received for analysis.The device returned with balloon folds expanded.Upon visual inspection of the device, a longitudinal tear was observed on the balloon, from the proximal cone, along the working length of the balloon.The balloon material was jagged and uneven at the tear site.Blood was visible in the inflation lumen.The distal shaft was flattened.The balloon failed negative prep.On pressurisation of the device, liquid was observed exiting the balloon working length.No other damage evident to the remainder of the device.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Additional information: annex d code added medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|