TORNIER INC TORNIER PERFORM REVERSED LATER ALIZED GLENOSPHERE +3 DIA 42MM; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Model Number DWJ024
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Fall (1848); Loss of Range of Motion (2032); Joint Dislocation (2374)
Date of Event 12/23/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device will not be returned.If additional information becomes available, it will be provided on a supplemental report.Device disposition unknown.
 
Event or Problem Description
It was reported that the patient fell on ice and got a dislocated shoulder.The emergency room was unable to relocate shoulder so the patient stayed overnight and then was taken to the operating room next morning for relocation under anesthesia.The patient remains pseudoparalytic and will be worked up for a revision rsa.
 
Event or Problem Description
It was reported that the patient fell on ice and got a dislocated shoulder.The emergency room was unable to relocate shoulder so the patient stayed overnight and then was taken to the operating room next morning for relocation under anesthesia.The patient remains pseudoparalytic and will be worked up for a revision rsa.
 
Additional Manufacturer Narrative
The reported event could not be confirmed.The device was not returned.Since the medical reports were provided, the opinion of the medical expert was requested and stated as following: "the previous surgeries and the trauma are enough reason for chronic instability and the need for (re-)revision of the right shoulder".More detailed information about the complaint event as well as the affected device must be available in order to determine the exact root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TORNIER PERFORM REVERSED LATER ALIZED GLENOSPHERE +3 DIA 42MM
Common Device Name
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
TORNIER INC
10801 nesbitt avenue s
bloomington MN 55437
Manufacturer (Section G)
TORNIER INC
10801 nesbitt avenue s
bloomington MN 55437
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key14206472
Report Number3004983210-2022-00055
Device Sequence Number4309747
Product Code KWS
UDI-Device Identifier00846832090396
UDI-Public00846832090396
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K161742
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date12/20/2024
Device Model NumberDWJ024
Device Catalogue NumberDWJ024
Device Lot NumberCZ1819347027
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/31/2022
Supplement Date Received by Manufacturer06/09/2022
Initial Report FDA Received Date04/26/2022
Supplement Report FDA Received Date07/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/20/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age29 YR
Patient SexMale
Patient Weight147 KG
-
-