• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAUSCH + LOMB STELLARIS VISION ENHANCEMENT SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAUSCH + LOMB STELLARIS VISION ENHANCEMENT SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number BL2330
Medical Device Problem Code Circuit Failure (1089)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/01/2022
Type of Reportable Event Serious Injury
Event or Problem Description
The user facility in the (b)(6) reported an error message appearing on the stellaris unit during surgery.Usm01 error, the ultrasound module was not detected in the system.The patient was transferred to another hospital for the surgery to be completed.
 
Additional Manufacturer Narrative
The field service technician on site started system check when usm01 error was displayed.The ultrasound module was replaced to eliminate error.An electrical safety test was performed, along with the main lead checked and secured.Final checks all within specification.Testing of the module found the main board was faulty, and the module was not rohs compliant.Field service records were reviewed, preventative maintenance was performed (b)(6) 2022 with no anomalies found.The device history was reviewed and found to meet manufacturing specification.This investigation ongoing.
 
Additional Manufacturer Narrative
The lot history, trend analysis, risk analysis and/or directions for use review were considered acceptable, with the product performing within anticipated rates.No corrective action required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STELLARIS VISION ENHANCEMENT SYSTEM
Common Device Name
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
BAUSCH + LOMB
rochester NY 14609
Manufacturer (Section G)
BAUSCH + LOMB, INC.
3365 treecourt industrial blvd
st. louis MO 63122
Manufacturer Contact
juli moore
MDR Report Key14207501
Report Number0001920664-2022-00055
Device Sequence Number19922914
Product Code HQC
UDI-Device Identifier00757770480678
UDI-Public0100757770480678
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K063331
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberBL2330
Device Lot Number222
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/21/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/01/2022
Supplement Date Received by Manufacturer05/25/2022
Initial Report FDA Received Date04/26/2022
Supplement Report FDA Received Date06/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/25/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-