• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR HI-TORQUE STEEL CORE 18 GUIDE WIRE WITH MICROGLIDE® COATING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR HI-TORQUE STEEL CORE 18 GUIDE WIRE WITH MICROGLIDE® COATING Back to Search Results
Model Number 1003282
Medical Device Problem Codes Break (1069); Difficult to Remove (1528); Failure to Advance (2524); Deformation Due to Compressive Stress (2889)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/04/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was received.Investigation is not yet complete.A follow up report will be submitted with all additional relevant information.The additional device is filed under a separate medwatch report number.
 
Event or Problem Description
It was reported that the procedure was performed to treat a lesion with heavily calcification and no tortuosity in the superficial femoral artery (sfa).The ht steelcore 18 guide wire was advanced but could not cross the lesion and during removal there was resistance with the anatomy and the tip became kinked and unraveled.A second guide wire was advanced, but had the same issue.Therefore, it was decided to use a non-abbott guide wire.There was no adverse patient effect and no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
A visual and dimensional inspection was performed on the returned guide wire and the reported core separation was confirmed.The reported failure to advance and remove were unable to be replicated in a testing environment as it was based on operational circumstances.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Based on the review of similar incident(s) there is no indication of a lot specific product quality issue.The investigation determined the reported failure to advance, difficulty to remove, tip break and kinks appear to be related to case circumstances of the procedure.In this case, based on the reported information, it is likely that during advancement and retraction against resistance, the coils likely stretched causing the appearance of kinks and a tip break.The noted bends on the core likely occurred during packing for return analysis.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HI-TORQUE STEEL CORE 18 GUIDE WIRE WITH MICROGLIDE® COATING
Common Device Name
GUIDE WIRE
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key14207598
Report Number2024168-2022-04491
Device Sequence Number1820735
Product Code DQX
UDI-Device Identifier08717648013959
UDI-Public08717648013959
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K982876
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2023
Device Model Number1003282
Device Catalogue Number1003282
Device Lot Number1120161
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/11/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/04/2022
Supplement Date Received by Manufacturer05/18/2022
Initial Report FDA Received Date04/26/2022
Supplement Report FDA Received Date05/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age52 YR
Patient SexMale
Patient Weight77 KG
-
-