CONCORD MANUFACTURING 2008T HEMODIALYSIS SYS., WITH CDX; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
|
Back to Search Results |
|
Model Number 190713 |
Device Problems
Thermal Decomposition of Device (1071); Melted (1385)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/21/2022 |
Event Type
malfunction
|
Event Description
|
A biomedical technician (biomed) at a user facility reported that a fresenius 2008t hemodialysis (hd) machine had a burned power switch and a melted wiring harness.The power switch appeared to have burn spots and a wire coming from it appeared melted and discolored.The burned parts were noticed during machine repair after the machine was initially removed from service for failing to power on.A patient was not connected to the machine at the time of the incident and there was no harm to any patients or individuals because of this malfunction.The machine has approximately 30,000 hours and the parts were the original fresenius part on the machine.The biomed reported that there was no damage observed on any other components, or any other additional issues, associated with the burned power switch and melted wiring harness.The biomed replaced the power switch and wiring harness, which resolved the issue.The unit was returned to service at the user facility without issue and without reoccurrence of the event.The parts were discarded by the biomed and are not available to be returned to the manufacturer for physical evaluation.
|
|
Manufacturer Narrative
|
Plant investigation: no parts were returned to the manufacturer for physical evaluation.Additionally, no on-site evaluation was performed by a fresenius field service technician (fst).However, the reported issue was confirmed based on the biomedical technician (biomed) description of the damage that was encountered during a machine evaluation.A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.
|
|
Search Alerts/Recalls
|
|
|