• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONCORD MANUFACTURING 2008T HEMODIALYSIS SYS., WITH CDX; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONCORD MANUFACTURING 2008T HEMODIALYSIS SYS., WITH CDX; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Model Number 190713
Device Problems Thermal Decomposition of Device (1071); Melted (1385)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/21/2022
Event Type  malfunction  
Event Description
A biomedical technician (biomed) at a user facility reported that a fresenius 2008t hemodialysis (hd) machine had a burned power switch and a melted wiring harness.The power switch appeared to have burn spots and a wire coming from it appeared melted and discolored.The burned parts were noticed during machine repair after the machine was initially removed from service for failing to power on.A patient was not connected to the machine at the time of the incident and there was no harm to any patients or individuals because of this malfunction.The machine has approximately 30,000 hours and the parts were the original fresenius part on the machine.The biomed reported that there was no damage observed on any other components, or any other additional issues, associated with the burned power switch and melted wiring harness.The biomed replaced the power switch and wiring harness, which resolved the issue.The unit was returned to service at the user facility without issue and without reoccurrence of the event.The parts were discarded by the biomed and are not available to be returned to the manufacturer for physical evaluation.
 
Manufacturer Narrative
Plant investigation: no parts were returned to the manufacturer for physical evaluation.Additionally, no on-site evaluation was performed by a fresenius field service technician (fst).However, the reported issue was confirmed based on the biomedical technician (biomed) description of the damage that was encountered during a machine evaluation.A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2008T HEMODIALYSIS SYS., WITH CDX
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer (Section G)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer Contact
jason busch
920 winter st
waltham, MA 02451
9043166958
MDR Report Key14207769
MDR Text Key290068545
Report Number0002937457-2022-00683
Device Sequence Number1
Product Code KDI
UDI-Device Identifier00840861100897
UDI-Public00840861100897
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K093902
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 04/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/26/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number190713
Device Catalogue Number190713
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeMO
Date Manufacturer Received04/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/30/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-