MAUDE Adverse Event Report: BIOMET MICROFIXATION TW DRILL 1.1X50MM 7MMSTOP W/NT; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET MICROFIXATION TW DRILL 1.1X50MM 7MMSTOP W/NT; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) Back to Search Results
Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/14/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported a male surgeon broke the tip of a drill while drilling a pilot hole during an open reduction and fixation of the maxilla.He recovered the broken tip and continued the operation.
 
Additional Manufacturer Narrative
Zimmer biomet complaint (b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.H6 component code: component code mechanical (g04) - drill item was returned for evaluation.Visual examination identified the tip of the device to be fractured.Scratches and other wear marks were noted.The part and lot information was verified.Review of the device history record(s) identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.
 
Event or Problem Description
No further event information is available at the time of this repot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TW DRILL 1.1X50MM 7MMSTOP W/NT
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
Manufacturer (Section D)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
Manufacturer (Section G)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
Manufacturer Contact
christina arnt
1520 tradeport drive
jacksonville, FL 32218
9047414400
MDR Report Key14207838
Report Number0001032347-2022-00122
Device Sequence Number1803276
Product Code HBE
UDI-Device Identifier00841036013165
UDI-Public(01)00841036013165(10)344288
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K062842
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number01-7144
Device Lot Number344288
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/18/2022
Supplement Date Received by Manufacturer07/13/2022
Initial Report FDA Received Date04/26/2022
Supplement Report FDA Received Date07/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/29/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexUnknown
-
-