| Model Number |
777624 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/30/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that after opened the package, the ureteral stent broke as the coil separated with the straight site.It was stated that a new stent had been unpacked to use.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event is confirmed, cause unknown.No physical sample was returned, however, a photo sample was submitted.A separation was noted on the proximal end of the stent.A potential root cause for this event could be, "inappropriate package design".As no patient involvement was reported the device was not used, however it is intended for treatment purposes.It is unknown if the device had met all relevant specifications.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "improper handling technique can seriously weaken the stent." "exercise care.Tearing of the stent can be caused by sharp instruments." "care should be exercised when removing the stent from inner polybag so as not to cause tearing or fragmentation." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
|
| |
|
Event or Problem Description
|
|
It was reported that after opened the package, the ureteral stent broke as the coil separated with the straight site.It was stated that a new stent had been unpacked to use.
|
| |
|