| Model Number |
08P1920 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/08/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow up report will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The customer reported a falsely decreased magnesium result generated on the alinity c analyzer on a patient.Results provided: (b)(6) 2022, sid (b)(6) = 1.9 / 1.1 / 1.8 mg/dl (normal range: 1.6-2.6 mg/dl).No impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of tickets determined that there is normal complaint activity for alinity c magnesium reagent lot number 48252un21.Trending review determined no trends for discrepant patient results for the product.Return testing was not completed as returns were not available.A review of the manufacturing documentation did not identify any issues associated with the customer's observation.A review of the product labeling concluded that the issue is sufficiently addressed.Based on the investigation, no systemic issue or deficiency of the alinity c magnesium reagent lot number 48252un21 was identified.
|
| |
|