| Catalog Number |
07027117190 |
| Medical Device Problem Codes |
High Test Results (2457); Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
The initial reporter received questionable elecsys cmv igg results for one patient tested on a cobas 8000 e 801 module serial number (b)(4).The patient had two samples collected.The patient's elecsys cmv igg results were reported outside the laboratory.The patient's samples were repeated on a vidas analyzer and a liaison analyzer.On (b)(6) 2022, the patient's first sample was collected.The patient's elecsys cmv igg result was 2.5 u/ml positive.The patient's vidas cmv igg result was 0.135 ua/ml negative.On (b)(6) 2022, the patient's second sample was collected.The patient's elecsys cmv igg result was "<1" u/ml negative.The patient's vidas cmv igg result was 19 ua/ml positive.The patient's liaison cmv igg result was 63 u/ml positive.
|
| |
|
Additional Manufacturer Narrative
|
|
The patient's sample was not available for investigation.Based on the provided information, the investigation did not identify a product problem.The cause of the event could not be determined.
|
| |
|