SYNTHES GMBH RING RETRACTOR DELTA-SHAPED Back to Search Results
Catalog Number 03.401.054
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/29/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Reporter is a j&j employee.Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2022, the ring retractor delta-shaped in question had been broken when the ring retractor was delivered to the hospital.The ring retractor was purchased last year.The hospital requested to return the broken ring retractor and deliver a new ring retractor.No further information is available.This report is for one (1) ring retractor delta-shaped this is report 1 of 1 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h6: the photo was returned to depuy synthes for evaluation.Visual analysis of the photo revealed that ring retractor delta-shaped has broken in the ring retractor.No other problems was reported.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for ring retractor delta-shaped.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h4, h6: part 03.401.054, lot 8724679: release to warehouse date: december 16, 2013.Manufacturing site: werk raron.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.H3, h6: a product investigation was completed: visual analysis of the returned sample revealed that the ring retractor delta-shaped has nicked in the ring retractor area.No other problem was reported.A dimensional inspection was not performed for the ring retractor delta-shaped due to the design of the device.The current and manufactured to drawings were reviewed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.¿the overall complaint was confirmed as the observed condition of the ring retractor delta-shaped would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RING RETRACTOR DELTA-SHAPED
Common Device Name
RETRACTOR
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK BALSTHAL (CH)
dornacherstrasse 20
balsthal 4710
SZ   4710
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key14208795
Report Number8030965-2022-02691
Device Sequence Number11217563
Product Code GAD
UDI-Device Identifier07611819352193
UDI-Public(01)07611819352193
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.401.054
Device Lot Number8724679
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/29/2022
Supplement Date Received by Manufacturer05/12/2022
06/29/2022
Initial Report FDA Received Date04/26/2022
Supplement Report FDA Received Date05/13/2022
07/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/16/2013
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-