Model Number 00-5901-021-00
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/04/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Customer has indicated that product will not be returned to zimmer biomet for investigation, as the product location is unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that during an initial knee procedure, a screwdriver fractured.The broken piece was retrieved and the surgery was completed with a second device without complication.No adverse events have been reported as a result of the malfunction.
 
Event or Problem Description
No further event information at time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; g3; h2; h3; h6; h10.Visual examination of the provided picture found the tip of the inserter fractured.The device was not returned for further evaluation.Device history record was reviewed and no discrepancies related to the reported event were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRILL PIN AND SCREW INSERTER
Common Device Name
EXTRACTOR
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key14208886
Report Number0001822565-2022-01193
Device Sequence Number1841541
Product Code HWB
UDI-Device Identifier00889024201637
UDI-Public(01)00889024201637(10)64027279
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number00-5901-021-00
Device Catalogue Number00590102100
Device Lot Number64027279
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/06/2022
Supplement Date Received by Manufacturer05/31/2022
Initial Report FDA Received Date04/26/2022
Supplement Report FDA Received Date06/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/15/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexMale
-
-