| Model Number |
FB-19C-1 |
| Medical Device Problem Codes |
Contamination (1120); Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/30/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is returned and an evaluation completed for it.The user¿s complaint was confirmed.Upon inspection and testing, it was observed that there was foreign material adhering to the link mechanism of the device.This was preventing the cup from opening.By ultrasonically cleaning the tip of the forceps, foreign matter was removed and the cup can be opened.Evaluation is ongoing.Supplemental report(s) will be submitted when any relevant new information is available.
|
| |
|
Event or Problem Description
|
|
As reported for this event by the customer, during pre-use inspection, the device cup did not open.There is no patient involvement and no harm reported to any patient.Another device was used for the intended procedure.Upon evaluation of the returned device, it was observed that there was foreign material adhering to the link mechanism of the device.This was preventing the cup from opening.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a definitive root cause could not be established.It is presumed that the adhesion of foreign matter is due to insufficient cleaning, or that it has not been sufficiently cleaned due to the passage of time after use.The instructions for use (ifu) states the following related to the suggested phenomenon: "·reprocess the instrument immediately after use by immersing it in a neutral, low-foaming, medical-grade detergent solution, then following the cleaning and sterilization procedures in this chapter.Failure to reprocess the instrument immediately after use, or using another type of detergent may cause corrosion at the instrument¿s distal end.This could cause components to fall off during use and may interfere with operation of the cups.·when inserting the instrument into the endoscope, hold it close to the biopsy valve or t-plug and keep it as straight as possible relative to the biopsy valve or t-plug.Otherwise, the insertion portion could be damaged.Lubrication 1.Immerse the insertion portion in the lubricant for 2 to 3 seconds.2.Remove the instrument from the lubricant.3.Operate the slider to open and close the cups two or three times.4.Wipe the exterior of the instrument with a clean, dry lint-free cloth and allow the instrument to air dry." olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|