• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ TORIC SINGLEPIECE IOL; LENS, INTRAOCULAR, TORIC OPTICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ TORIC SINGLEPIECE IOL; LENS, INTRAOCULAR, TORIC OPTICS Back to Search Results
Model Number SN6AT9
Medical Device Problem Codes Device Dislodged or Dislocated (2923); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Codes Blurred Vision (2137); Visual Impairment (2138)
Date of Event 08/18/2021
Type of Reportable Event Serious Injury
Event or Problem Description
A non healthcare professional reported that following implantation of an intraocular lens (iol) the lens was dislocated due to peripheral capsulotomy.The patient experienced double vision.The lens was explanted in a secondary procedure.Additional information was requested.
 
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
The lens was returned wrapped in gauze inside a small brown envelope labeled miller.The lens was adhered to a small piece of paper with viscoelastic.The lens was cleaned with klrp.The lens was damaged on the edge.Power and resolution were verified.The optical resolution was acceptable; lens met specification readings for an company lens focal length 25.0 and cylinder 6.00.The root cause for the reported visual issues could not be determined.No problem was found with the returned lens.Each lens is subjected to a 100% assessment of the power (spherical and cylinder) and optical resolution during the manufacturing process in order to determine acceptability per the lens model and diopter.Information was provided that the toric lens (6.00cyl), 25.0 diopter lens was exchanged for a non-toric 21.0 diopter lens.The difference of 6.00 cylinder power would indicate a calculation issue may have occurred.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
Additional information was requested and received stating the patient experienced blurred visual acuity.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ TORIC SINGLEPIECE IOL
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14209524
Report Number1119421-2022-00858
Device Sequence Number4383344
Product Code MJP
UDI-Device Identifier00380652279192
UDI-Public00380652279192
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/29/2025
Device Model NumberSN6AT9
Device Catalogue NumberSN6AT9.250
Device Lot Number15117954
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/04/2022
Supplement Date Received by Manufacturer04/25/2022
Initial Report FDA Received Date04/26/2022
Supplement Report FDA Received Date05/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/30/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MONARCH III IOL CARTRIDGE D.; MONARCH III IOL INJECTOR.; PROVISC.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age67 YR
Patient SexMale
-
-