Model Number HB205-10
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/31/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device history record (dhr) review was conducted for the reported lot/serial number.The device was released meeting all qa specifications.We are currently unable to establish a relationship between the device and the issue reported.Therefore visual and functional analysis of the product could not be performed.We are unable to confirm a relationship between the device and the issue reported and a definitive root cause for the reported event could not be determined.The information obtained during complaint investigation will be included in our global complaint trending and product surveillance will continue to monitor complaints of this type for adverse trends.If the product is received or additional information is obtained, the investigation will be reopened accordingly per standard operating procedure.
 
Event or Problem Description
It was reported that during a localizer procedure on (b)(6) 2022, the tag broke sometime between the placement and the surgery.No injury to the patient was reported, the device broke sometime after the tag placement, the mammography after the placed looked normal.The procedure was completed and no additional procedures required at the moment.An image provided could see the tag broken on the specimen removed.No other information at the moment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCALIZER
Common Device Name
MARKER, RADIOGRAPHIC, IMPLANTABLE
Manufacturer (Section D)
HEALTH BACONS, INC
250 campus drive
marlborough MA 01752
Manufacturer (Section G)
HOLOGIC, INC.
250 campus drive
marlborough, MA 01752
Manufacturer Contact
daniel guevara
562 parkway
coyol free zone building b24
san jose, alajuela 20102- CRI
CS   20102 CRI
MDR Report Key14209529
Report Number3013649990-2022-00010
Device Sequence Number17238999
Product Code NEU
UDI-Device Identifier00855436007044
UDI-Public00855436007044
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K193189
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/27/2026
Device Model NumberHB205-10
Device Catalogue NumberHB205-10
Device Lot Number48461
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/31/2022
Initial Report FDA Received Date04/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/27/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-