• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PERFUSION SYSTEMS BIO-MEDICUS NG ARTERIAL CANNULA; CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PERFUSION SYSTEMS BIO-MEDICUS NG ARTERIAL CANNULA; CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 96570-117
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/29/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information that during use of a bio-medicus nextgen cannula, it was reported a that the connector was cracked.The device use was unspecified.There was no patient impact associated with this event.
 
Additional Manufacturer Narrative
Product analysis: visual inspection showed evidence of damage/cracks in the connector.The reason for return was confirmed.Conclusion: after investigation, the complaint is confirmed for a cracked cannula body connector.It is unknown what may have caused this occurrence, however, based on the available information, this complaint is potentially the result of a use-related issue.The damage to the cannula connector can occur when the introducer is inserted into the cannula body without the use of the hemostasis cap.The introducer handle can cause stretching, crazing, or cracking when it is forced into the connector.The hemostasis cap is clear and is attached to the introducer on the provided protective sheath.From the photo provided by the customer, the introducer is in fact inserted into the cannula body without the hemostasis cap, confirming that this is most likely a use-related issue.Trends for issues with this device are monitored.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIO-MEDICUS NG ARTERIAL CANNULA
Common Device Name
CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer (Section G)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key14209745
Report Number2184009-2022-00094
Device Sequence Number13904275
Product Code DWF
UDI-Device Identifier00643169025776
UDI-Public00643169025776
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K142673
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/08/2025
Device Model Number96570-117
Device Catalogue Number96570-117
Device Lot Number222539682
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/30/2022
Supplement Date Received by Manufacturer06/22/2022
Initial Report FDA Received Date04/26/2022
Supplement Report FDA Received Date07/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-