• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HEMOLOK XL CLIPS 6/CART 84/BOX; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HEMOLOK XL CLIPS 6/CART 84/BOX; CLIP, IMPLANTABLE Back to Search Results
Model Number IPN915186
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/14/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device history review could not be conducted since the lot number was not provided.
 
Event or Problem Description
When the user ligated a clip during a robot-assisted laparoscopic nephrectomy, it got broken at the hook.The broken hook fell in the patient, but was removed by suction.The user opened another cartridge to complete the surgery.
 
Additional Manufacturer Narrative
(b)(4).Although the lot number was not reported, a potential lot number was found by looking at the sales history.The potential lot number is 73g2100704.Per dhr the product hemolok xl clips 6/cart 84/box lot# 73g2100704 was manufactured on 07/26/2021 a total of (b)(4) pieces.Lot was released on 08/05/2021.Dhr investigation did not show issues related to complaint.The customer returned one clip from unit 544250 hemolok xl clips 6/cart 42/box for investigation.The cartridge was not returned.The loose clip was visually examined with and without magnification.Visual examination revealed that the clip was broken on its hook leg (broken at hook side of hook bosses) and also had both pierced bosses broken off.The broken pieces were not returned.Biological material was observed on the clip, indicating that the sample was used.R & d engineering was consulted for this complaint issue.According to r & d, the observed damage near the hook is consistent with improper loading of the clip or with using damaged appliers.The appliers were not returned for investigation.The ifu for this product, l02425, was reviewed as a part of this complaint investigation.The ifu states, "always check the alignment of the applier jaws before use.When closed, jaw tips should be directly aligned and not offset.Alignment of the jaw is critical for safe application of the clip.If this is not done, patient injury may occur.Proper maintenance, care and cleaning are necessary to ensure proper functionality." a corrective action is not required at this time as the damage observed on the clip indicates that unintentional user error caused or contributed to this event.The reported complaint of "broken clip - multiple locations" was confirmed based upon the sample received.One loose clip was returned with its hook leg broken at the hook side of the hook bosses and also with both pierced bosses broken off.The broken pieces were not returned.R & d was consulted for this complaint and it was determined that the observed damage to the broken clip is consistent with improper loading of the clip or with using damaged appliers.It is unknown how the returned clip was handled during use and the appliers were not returned for investigation.Therefore, based upon the damage observed, unintentional user error caused or contributed to this event.
 
Event or Problem Description
When the user ligated a clip during a robot-assisted laparoscopic nephrectomy, it got broken at the hook.The broken hook fell in the patient, but was removed by suction.The user opened another cartridge to complete the surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMOLOK XL CLIPS 6/CART 84/BOX
Common Device Name
CLIP, IMPLANTABLE
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
rancho el descanso
tecate 21478
MX   21478
Manufacturer Contact
jasmine brown
3015 carrington mill blvd
morrisville, NC 27560
9193614124
MDR Report Key14211718
Report Number3003898360-2022-00148
Device Sequence Number19283429
Product Code FZP
UDI-Device Identifier24026704695953
UDI-Public24026704695953
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K133202
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIPN915186
Device Catalogue Number544250
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/15/2022
Supplement Date Received by Manufacturer06/03/2022
Initial Report FDA Received Date04/26/2022
Supplement Report FDA Received Date06/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-