Information contained in this report was previously submitted through asr/psr on 15-aug-2018.A review of the device history record has been completed.No deviations or non-conformance's noted.Device evaluation: visual analysis of the returned device identified: opening, weight to the spec, yellow particles.A microscopic analysis was performed which identify a striated edge opening, consist with use of a surgical tool.Based on the device analysis the final assessment is: a striated edge opening on anterior side due to surgical damage.During the analysis was observe along the same opening striated pattern.The event of "capsular contracture" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade iii, concern for textured implant device, "breast implant illness, personal injuries".
|