Model Number 1550250-28
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Dyspnea (1816); Stenosis (2263); Swelling/ Edema (4577)
Date of Event 04/04/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.The reported patient effects of stenosis and edema are listed in the xience sierra everolimus eluting coronary stent systems instructions for use (ifu) as known patient effects of coronary stenting procedures.A conclusive cause for the reported patient effects, and the relationship to the product, if any, cannot be determined; however, the subsequent treatment appears to be related to the operational context of the procedure.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event or Problem Description
Patient id: (b)(6).It was reported that the procedure was performed on 10/02/2020 to treat a lesion in the proximal left anterior descending (lad) artery.A 2.75 x 15 mm xience sierra stent and a 2.5 x 28 mm xience sierra stent were implanted.On 04/04/2022, the patient was re-hospitalized with shortness of breath and lower extremity edema.Coronary catheterization was performed and noted 80% in-stent restenosis in the proximal and mid lad, in the 2.5 x 28 mm xience sierra stent and a previously implanted unspecified stent, and 90% stenosis in the circumflex (cx) artery.Angioplasty was performed to treat the in-stent restenosis in the 2.5 x 28 mm xience sierra stent.The event resolved on (b)(6) 2022.A staged procedure will be performed at a later date to treat the cx artery.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key14212378
Report Number2024168-2022-04504
Device Sequence Number4170724
Product Code NIQ
UDI-Device Identifier08717648227127
UDI-Public08717648227127
Combination Product (Y/N)Y
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/05/2022
Device Model Number1550250-28
Device Catalogue Number1550250-28
Device Lot Number0061841
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/05/2022
Initial Report FDA Received Date04/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/06/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age73 YR
Patient SexFemale
Patient Weight60 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-