It was reported that, after thr surgery had been performed (b)(6) 2022, it was noticed that the liner was never seated in shell, causing the head and cup to rub.This adverse event was treated by revision surgery on (b)(6) 2022 current health status of patient is unknown.The femoral head presented significant wear.
|
|
H3, h6: the device was not returned for evaluation.The pictures provided were reviewed and could not confirm the stated failure mode.The clinical/medical investigation concluded that, the provided photos were reviewed but do not aid in confirming the clinical root cause of the reported liner never seated in shell.Smith and nephew has not received the device/ adequate materials to fully evaluate the complaint, but if additional clinically relevant materials are later received, then the case may be re-opened for further evaluation.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A review of the instructions for use for this device loosening event has been identified in possible adverse effects section as a result from trauma, improper implant selection, improper implant positioning, improper fixation, and/or migration of the components.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but not limited to surgical technique used, size of device or user/procedural variance.The contribution of the device to the reported event could not be corroborated, however, the pictures do confirm that the head is worn.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|