| Model Number |
24702 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/20/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The patient presented with peripheral artery disease (pad).The 90% stenosed target lesion was located in the mildly tortuous and severely calcified superficial femoral artery.A 2mm x 40mm x 145cm coyote es balloon catheter was advanced for dilation.However, during initial inflation at 12 atmospheres for 10 seconds, the balloon ruptured on the middle part.The device was removed.A 2.0mm x 100mm x 90cm sterling sl balloon catheter was advanced for dilatation.The balloon also ruptured during initial inflation at 12 atmospheres for 10 seconds.The device was removed and the procedure was completed with a different device.There were no patient complications nor injuries reported and the patient was in good condition.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The patient presented with peripheral artery disease (pad).The 90% stenosed target lesion was located in the mildly tortuous and severely calcified superficial femoral artery.A 2mm x 40mm x 145cm coyote es balloon catheter was advanced for dilation.However, during initial inflation at 12 atmospheres for 10 seconds, the balloon ruptured on the middle part.The device was removed.A 2.0mm x 100mm x 90cm sterling sl balloon catheter was advanced for dilatation.The balloon also ruptured during initial inflation at 12 atmospheres for 10 seconds.The device was removed and the procedure was completed with a different device.There were no patient complications nor injuries reported and the patient was in good condition.It was further reported that both balloon catheters were inflated twice before they ruptured.
|
| |
|
Search Alerts/Recalls
|