|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).The expiration date is currently unavailable.The complaint device is not being returned, therefore unavailable for a physical evaluation, however a photo was provided.Upon visual inspection of the photo, it could be observed the two devices along with their anchors.Both device's shafts are in good condition.It is not possible to verify the breakages on the anchors due to the photos are blurry and not detailed.A manufacturing record evaluation was performed for the finished device 6l65877 number, and no non-conformances were identified.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.According with the visual inspection of the photo, this complaint cannot be confirmed.The devices are required for evaluation, the photos provided does not contain enough evidence to determine why the customer experienced the failure, hands on analysis should provide the required evidence to provide a root cause.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
|