The patient's dob, weight, ethnicity, and race are unknown.This information was not available from the facility.Patient information regarding relevant tests/laboratory data is unknown.This information was not available from the facility.Report source: foreign- taiwan.Visual inspection found a damaged distal tip and a damaged distal bond with one leaflet lifted, but remained intact.During functional testing, using an indeflator filled with green color dye water, the device was pressurized and at less than 2 atm, a longitudinal balloon tear was observed.The ifu warns at least 99.9% of the balloons (with a 95% confidence level) will not burst at or below the rbp.Additionally, retained device component is listed as possible adverse effect of the procedure.
|
|
The angiosculpt device was used to treat a severely calcified mid lad.During inflation at 16 atm, the balloon ruptured.The procedure was completed with a different device.No patient injury reported.During the device evaluation, the balloon rupture was confirmed and a distal bond peel was observed.This product problem is being submitted conservatively because the balloon ruptured below rbp and the distal bond peeled.
|