• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THE SPECTRANETICS CORPORATION ANGIOSCULPT RX PTCA; CATHETER, PTCA, CUTTING/SCORING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THE SPECTRANETICS CORPORATION ANGIOSCULPT RX PTCA; CATHETER, PTCA, CUTTING/SCORING Back to Search Results
Model Number 2200-3010
Medical Device Problem Codes Peeled/Delaminated (1454); Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/30/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The patient's dob, weight, ethnicity, and race are unknown.This information was not available from the facility.Patient information regarding relevant tests/laboratory data is unknown.This information was not available from the facility.Report source: foreign- taiwan.Visual inspection found a damaged distal tip and a damaged distal bond with one leaflet lifted, but remained intact.During functional testing, using an indeflator filled with green color dye water, the device was pressurized and at less than 2 atm, a longitudinal balloon tear was observed.The ifu warns at least 99.9% of the balloons (with a 95% confidence level) will not burst at or below the rbp.Additionally, retained device component is listed as possible adverse effect of the procedure.
 
Event or Problem Description
The angiosculpt device was used to treat a severely calcified mid lad.During inflation at 16 atm, the balloon ruptured.The procedure was completed with a different device.No patient injury reported.During the device evaluation, the balloon rupture was confirmed and a distal bond peel was observed.This product problem is being submitted conservatively because the balloon ruptured below rbp and the distal bond peeled.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGIOSCULPT RX PTCA
Common Device Name
CATHETER, PTCA, CUTTING/SCORING
Manufacturer (Section D)
THE SPECTRANETICS CORPORATION
5055 brandin court
fremont CA 94538
Manufacturer (Section G)
SPECTRANETICS
5055 brandin court
fremont CA 94538
Manufacturer Contact
diana melliza galvez
5055 brandin court
fremont, CA 94538
MDR Report Key14213657
Report Number3005462046-2022-00017
Device Sequence Number16595117
Product Code NWX
UDI-Device Identifier00813132021207
UDI-Public00813132021207
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
P050018
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/16/2023
Device Model Number2200-3010
Device Catalogue Number2200-3010
Device Lot NumberG20010019
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/01/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/30/2022
Initial Report FDA Received Date04/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/16/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ASAHI: 0.014" SION GUIDEWIRE; MEDTRONIC: 6F LAUNCHER GUIDE CATHETER
Patient Age64 YR
Patient SexMale
-
-