• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON AND CO. BD NANO¿ ULTRA-FINE¿ PEN NEEDLE; HYPODERMIC SINGLE LUMEN NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON AND CO. BD NANO¿ ULTRA-FINE¿ PEN NEEDLE; HYPODERMIC SINGLE LUMEN NEEDLE Back to Search Results
Model Number 8975
Medical Device Problem Code Complete Blockage (1094)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/20/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Fda notified: the initial reporter notified the fda via uf/importer report#: mw5108043.Investigation summary: no samples were returned therefore the complaint could not be confirmed and the root cause is undetermined.This is the 2nd related complaint for the reported lot number.A review of the manufacturing records was performed and no non-conformance's were raised in association with this type of event for this lot.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.A complaint history check was performed and this is the 2nd.Related complaint for needle clog on lot # 1103587.A lot history review was carried out and no related non conformance's were raised in association with this packaged lot concluding all inspections were performed as per the applicable operations and met qc specifications.As no samples and/or photo(s) were received the investigation concluded: unconfirmed: embecta was not able to duplicate or confirm the customer¿s indicated failure as no samples or photos were returned.Root cause cannot be determined at this time as the issue is unconfirmed as no samples or photos were returned.Based on the above, no additional investigation and no capa/sa is required at this time.
 
Event or Problem Description
It was reported that the bd nano¿ ultra-fine¿ pen needle was unable to deliver medication.The following information was provided by the initial reporter: defective needles.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD NANO¿ ULTRA-FINE¿ PEN NEEDLE
Common Device Name
HYPODERMIC SINGLE LUMEN NEEDLE
Manufacturer (Section D)
BECTON DICKINSON AND CO.
pottery road
dun laoghaire co
Manufacturer (Section G)
BECTON DICKINSON AND CO.
pottery road
dun laoghaire co
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key14214874
Report Number9616656-2022-00457
Device Sequence Number11217484
Product Code FMI
UDI-Device Identifier10885403897511
UDI-Public10885403897511
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K133532
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/30/2026
Device Model Number8975
Device Catalogue Number320883
Device Lot Number1103587
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/01/2022
Initial Report FDA Received Date04/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/13/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-