| Model Number |
SSLP-400 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
11/15/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture.
|
| |
|
Event or Problem Description
|
|
Company representative reported on behalf of the physician, right side capsular contracture baker grade iii/iv.The device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified, weight to the spec, crease fold.Cloudy was observed, after autoclave disinfection process.Based on the device analysis, the final assessment is: no issues found related with the manufacturing process.
|
| |
|
Event or Problem Description
|
|
Company representative reported, on behalf of the physician.Right side capsular contracture, baker grade iii/iv.The device has been explanted.
|
| |
|
Search Alerts/Recalls
|