• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION MUSTANG; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION MUSTANG; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number 24674
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/09/2022
Type of Reportable Event Malfunction
Event or Problem Description
Reportable based on device analysis completed on (b)(6) 2022.It was reported that crossing difficulties were encountered.The more than 70% stenosed target lesion was located in an arteriovenous fistula in a severely tortuous and severely calcified vessel.A 8.0 x 40, 135cm mustang balloon catheter was advanced for dilatation but failed to cross the lesion.The procedure was completed with another of same device.There were no patient complications reported and the patient condition was good.However, returned device analysis revealed balloon longitudinal tear.
 
Additional Manufacturer Narrative
A mustang 8.0 x 40 was returned for analysis.A visual examination of the returned device found that the balloon had been inflated and was not refolded.It is not known when the balloon was subjected to positive pressure.A balloon longitudinal tear was identified beginning approximately 4mm proximal of the proximal markerband and extending approximately 26mm distally across the balloon material.An examination of the balloon material and markerbands identified no issues which could potentially have contributed to this complaint.A visual examination observed damage to the tip of the device.This type of damage is consistent with the device encountering resistance and excessive force being applied when attempting to cross a lesion.A visual and tactile examination found no kinks or damage to the shaft of the device.A visual examination observed no issues with the markerbands of the device.Both markerbands were undamaged and present on the shaft of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MUSTANG
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key14214992
Report Number2134265-2022-04831
Device Sequence Number19882351
Product Code FGE
UDI-Device Identifier08714729793663
UDI-Public08714729793663
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K141521
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/03/2023
Device Model Number24674
Device Catalogue Number24674
Device Lot Number0026457570
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/06/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/21/2022
Initial Report FDA Received Date04/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/03/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age77 YR
Patient SexFemale
-
-