This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: tip broken.Probable root cause: non-conforming component, poor assembly process, misuse.Use error the reported failure mode will be monitored for future reoccurrence.Note: the investigation was previously closed based on product not received; however, the product has now been physically received at stryker endoscopy, usa and the investigation has been re-opened.Investigation as follows is now based on product received.Alleged failure: the tip of the probe was broken off.It was unknown whether the broken tip could be retrieved or not.The failure(s) identified in the investigation is consistent with the complaint record.The probable root cause/s could be excessive force used when inserting removing the probe, a probe inserted into an obstructed passageway or any forceful impact to the tip of the probe with rigid objects.The probe is used as a tool for a mechanical displacement of tissue or bone, or to pry or scrub.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|