CORDIS CORPORATION TRAPEASE PVCF FEM/JUG 55CM CSI; FILTER, INTRAVASCULAR, CARDIOVASCULAR
|
Back to Search Results |
|
| Model Number |
466P306AU |
| Health Effect - Clinical Codes |
Perforation of Vessels (2135); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
09/01/2021
|
|
Type of Reportable Event
|
Death
|
|
Additional Manufacturer Narrative
|
|
Occupation: other, senior counsel, litigation.Please note that the exact event date is unknown and the event date is the complaint awareness date.(b)(4).As reported a patient underwent placement of a trapease vena cava filter.The information provided indicated that the filter subsequently caused perforation of the filter.The information also indicated that the patient is deceased, however a cause of death and relationship to the filter was not provided.Additional information indicated that the patient became aware of perforation of filter strut(s) outside the inferior vena cava (ivc).Approximately fourteen years and three months post implant.It was reported that the patient also experienced consistent pain on the left leg and groin area.According to the medical records the indication for the filter implant was pulmonary emboli.The filter was implanted via the right femoral vein and deployed into the infrarenal area of the inferior vena cava (ivc) with the leading edge of the filter at the l2 level (right pedicle).The patient tolerated the procedure well without evidence of complications.The product was not returned for analysis.A review of the device history record revealed no anomalies during the manufacturing and inspection processes that can be associated with the reported complaint.The trapease vena cava filter is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the vena cava for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pe where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pe where anticoagulant therapy has failed, or is contraindicated.The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart.Vessel perforation is a known adverse event associated with implanting vena cava filters and is listed as such in the instructions for use (ifu) and notes vessel damage such as intimal tears and perforation as procedural and long-term complications related to ivc filters.Without procedural films or post implant imaging available for review, the reported events could not be confirmed or further clarified.It was noted in the information provided that the patient is deceased, with the limited information provided and no cause of death reported a clinical determination or relationship to the filter could not be made.Pain does not represent a device malfunction and may be related to underlying patient specific issues.Given the limited information available for review at this time, there is nothing to suggest that the reported events are related to the design and manufacturing process of the device; therefore, no corrective action will be taken.Should additional information become available, the file will be updated accordingly.Please note that this is the initial/final report for this product.
|
| |
|
Event or Problem Description
|
|
As reported by the legal brief, the patient underwent placement of a trapease vena cava filter.The report states that the filter subsequently malfunctioned and caused injury and damage to the patient including, but not limited to perforation of the filter.The patient is deceased, but there was no information provided to relate his demise to the device.As a direct and proximate result of these malfunctions, the patient suffered life-threatening injuries and damages, and required extensive medical care and treatment.As a further proximate result, the patient has suffered significant medical expenses pain and suffering, loss of enjoyment of life and other damages.Information received per the medical records indicate that the patient has a history of pulmonary emboli.The filter was implanted via the patient's right femoral vein using the seldinger technique.Contrast was injected to confirm there were no clots within the femoral or iliac veins.A venocavogram was performed.This demonstrated the origin of the renal veins and the caval diameter (3 cm).The filter was placed into the infrarenal area of the inferior vena cava (ivc) with the leading edge of the filter at the l2 level (right pedicle).The patient tolerated the procedure well without evidence of complications.Information received per the patient profile form (ppf) states that the patient experienced perforation of filter strut(s) outside the ivc.The patient became aware of the reported events approximately fourteen years and three months after the index procedure.The patient also experienced pain consistently (left leg and groin area).
|
| |
|
Search Alerts/Recalls
|
|
|