| Model Number |
MICL12.6 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Intraocular Pressure Increased (1937); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
03/04/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The reporter indicated that the surgeon implanted a 12.6mm micl12.6; -11.5 diopter implantable collamer lens into the patient's right eye (od) on (b)(6) 2022.The patient experienced angle closure with elevated intraocular pressure.On (b)(6) 2022 the lens was exchanged with a shorter length lens and this resolved the problem.The cause of the event was reported as the device.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Type of investigation: 4110 - lens work order search: no similar complaint type events reported for units within the same lot.Claim#: (b)(4).
|
| |
|