MAUDE Adverse Event Report: ALCON GRIESHABER AG DIATHERMY PROBE DSP; UNIT, PHACOFRAGMENTATION, ACCESSORY, PROBE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON GRIESHABER AG DIATHERMY PROBE DSP; UNIT, PHACOFRAGMENTATION, ACCESSORY, PROBE Back to Search Results
Catalog Number 339.31
Medical Device Problem Code Coagulation in Device or Device Ingredient (1096)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/11/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A physician reported that a diathermy probe exhibited no electrocoagulation function before the vitrectomy surgery.The surgery was completed after a new replacement.There was no patient harm.
 
Additional Manufacturer Narrative
The investigation is completed.A review of the device history record (dhr) traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The sample is received without inner blister, outer blister or cover foil.The sample shows no macroscopic signs of damage.The instrument was visually inspected.No abnormality was observed.Electrical resistance measurements indicate insufficient contact.When the instrument was opened, the contacts were not destroyed.Nevertheless, no defect contact points in the instrument could be identified.All contact points were specifically investigated but non of the contact points was loose.Based on the outcome of this investigation the complaint of the customer can be confirmed but the root cause could not be determined anymore.In production a 100% inspection of the instrument is performed which would identify defect contact points.Since the instrument passed this test, the contact must have been affected later.But this cannot be determined anymore.The root cause cannot be identified conclusively because the sample has pass the 100% inspection.As the blister was opened by the customer, he handled the product.The point in time when the malfunction occurred, can no longer be determent.This complaint has been reviewed and future data will be monitored for evidence of adverse trending and further action will be taken, as appropriate.At a minimum, this will include completing reviews of complaint class report levels on a monthly basis.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIATHERMY PROBE DSP
Common Device Name
UNIT, PHACOFRAGMENTATION, ACCESSORY, PROBE
Manufacturer (Section D)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ  8203
Manufacturer (Section G)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ   8203
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14216592
Report Number3003398873-2022-00025
Device Sequence Number11796699
Product Code HQC
UDI-Device Identifier07612717071384
UDI-Public07612717071384
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K063583
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2022
Device Catalogue Number339.31
Device Lot NumberF175425
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/31/2022
Supplement Date Received by Manufacturer06/30/2022
Initial Report FDA Received Date04/27/2022
Supplement Report FDA Received Date07/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/21/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-