| Model Number |
140-9800 |
| Medical Device Problem Code |
Failure to Disconnect (2541)
|
| Health Effect - Clinical Code |
Cerebrospinal Fluid Leakage (1772)
|
| Date of Event |
04/01/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the during a superion indirect decompression spacer implant procedure, the device malfunctioned and the dilator advanced too far anterior.The patient experienced a dura injury at l4-l5 of the lumbar vertebrae and had a cerebrospinal fluid (csf) leak.The physician administered a blood patch and the implant procedure was completed successfully.The patient is doing well post operatively.
|
| |
|
Event or Problem Description
|
|
It was reported that the during a superion indirect decompression spacer implant procedure, the device malfunctioned and the dilator advanced too far anterior.The patient experienced a dura injury at l4-l5 of the lumbar vertebrae and had a cerebrospinal fluid (csf) leak.The physician administered a blood patch and the implant procedure was completed successfully.The patient is doing well post operatively.
|
| |
|
Additional Manufacturer Narrative
|
|
The complaint was confirmed.The cannula assembly could not be inserted over the dilator due to the presence of metallic protrusions on the inner surface area of the dilator 2.A welding process that was performed in error resulted in a weld intrusion.In addition, the reported allegation of a csf leak is a known inherent risk with the use of the device.
|
| |
|