| Model Number |
20210 |
| Medical Device Problem Code |
Nonstandard Device (1420)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Intended use: api® 20 c aux is a qualitative standardized system for the precise identification of the most frequently encountered yeasts.It uses miniaturized tests as well as a specially adapted database.Inoculation and reading of the strip are performed manually and the identification is obtained using an identification software.Incident description: a customer in (b)(6) reported to biomérieux a strip issue leading to "no result" in association with the product api 20 c aux 25 strips+25media (ref 20210, batch number 1009159121, expiry date 03-jul-2023) with four patient's strains (ophthalmic samples).The customer reported an issue with one batch of api 20c strips where the cups dried out too quickly.It was reported that the lid didn't close tightly because it was deformed, but even when adding tape it dried out and the water put under the strip evaporated completely.The customer performed an antifungiogram without having the identification because no other methods were available.Per photos sent by the customer, the water in the honey-wells and the suspension inside the strips wells are dehydrated.Consequently, the profile cannot be read.It was highlighted by biomérieux global customer service that identification is mandatory to interpret the susceptibility testing results as breakpoints may differ by species.There is no indication or report from the customer that this event led to any adverse event related to the patient's state of health.An investigation has been initiated.
|
| |
|
Additional Manufacturer Narrative
|
|
An internal investigation was performed following notification from a customer in france of an issue they were having with a batch of api 20 c aux 25 strips+25media (ref# 20210).No results were obtained with patient samples and it was noted that the honey-well were dry, thus results could not be obtained.Investigation: product submitted by the customer and the retained sample for the impacted lot number (1009159121) were tested in parallel with one internal lot number used as reference.A visual verification of the incubation boxes and strips was performed.No issue was observed on the product returned and on the reference lot.Three incubation boxes showed a slight deformation on the retained samples of lot number 1009159121.Tests were performed by inoculating api 20 c aux strips with c medium (without specimen) and incubating at 30°c.We included in the testing the three incubation boxes showing a slightly deformation.No dehydration of cupules was observed up to 48h of incubation at 30°c on all the strips tested.Batch record analysis.No non-conformity during manufacturing and quality control processes were observed for the lot 1009159121 which might explain the issue.Complaint analysis: as of the date of the investigation, no other customers have reported any similar issues on against the customer's lot.Conclusion: the investigation did not identify any product performance issue, thus the root cause remains unknown.The results obtained during our investigation on the impacted api 20 c aux lot (reference 20210 lot 1009159121) did not show issues related to cupules dehydration.The batch record analysis did not show non-conformity during manufacturing and quality control processes for the lot number 1009159121.Analysis of the complaints does not show any deviation on the product reference api 20 c aux (reference 20210).
|
| |
|
Search Alerts/Recalls
|