| Model Number |
24672 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/20/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The target lesion was located in an arteriovenous fistula.A 7.0 x 120, 75cm mustang balloon catheter was advanced for dilatation.However, during inflation, the balloon ruptured.The procedure was completed with another of the same device.No patient complications were reported and the patient status was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: a mustang device was returned for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.The returned device was attached to an encore inflation unit.Positive pressure was applied and no leaks or issues was noted with the balloon material.A visual examination of the marker bands identified no issues which may potentially have contributed to the complaint incident.A visual and tactile examination identified no kinks or damage to the shaft which may have potentially contributed to the complaint incident.A visual and tactile examination identified no damage to the tip which may have potentially contributed to the complaint incident.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The target lesion was located in an arteriovenous fistula.A 7.0 x 120, 75cm mustang balloon catheter was advanced for dilatation.However, during inflation, the balloon ruptured.The procedure was completed with another of the same device.No patient complications were reported and the patient status was stable.
|
| |
|
Search Alerts/Recalls
|