Model Number 8884720221
Medical Device Problem Code Deflation Problem (1149)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 03/15/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event or Problem Description
The customer reported they were treating a 16 year old female patient with a diagnosis of spastic cerebral palsy.They were having a problem with the balloon which was damaged.The device is removed.
 
Additional Manufacturer Narrative
The device history record (dhr) was reviewed, and no abnormal process conditions were present during the manufacturing of the product that could have led to the reported condition.The dhr review showed that all acceptance criteria inspections per established sampling levels were within acceptable limits during the production process.A sample analysis could not be performed because no photo or sample was available for evaluation.The complaint will be reopened if a sample is received.The reported condition could not be confirmed.The investigation was carried out with the multifunctional team.All processes and controls were found to be followed correctly, including packaging and all inspections performed on the product.No abnormal conditions were found that could trigger the reported condition.A risk assessment was carried out to determine the observed occurrence for this failure mode.The observed occurrence is below the range of the expected occurrence; therefore, no action plan is required at this point.This complaint will be used for tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GSTRO FEED TBE W/Y PRT 22FR
Common Device Name
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
CARDINAL HEALTH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX  22500
Manufacturer (Section G)
CARDINAL HEALTH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX   22500
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key14218435
Report Number9612030-2022-03233
Device Sequence Number17490064
Product Code KNT
UDI-Device Identifier10884521007321
UDI-Public10884521007321
Combination Product (Y/N)N
Initial Reporter CountryCO
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number8884720221
Device Catalogue Number8884720221
Device Lot Number2115220264
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/12/2022
Supplement Date Received by Manufacturer04/12/2022
Initial Report FDA Received Date04/27/2022
Supplement Report FDA Received Date05/23/2022
Usage of Device A
Patient Sequence Number1
Patient Age16 YR
Patient SexFemale
-
-