• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. BEARING +3 MM THICKNESS 40 MM DIAMETER; SHOULDER PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. BEARING +3 MM THICKNESS 40 MM DIAMETER; SHOULDER PROSTHESIS Back to Search Results
Catalog Number 110031428
Medical Device Problem Codes Detachment of Device or Device Component (2907); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Failure of Implant (1924); Muscle Weakness/Atrophy (1967); Joint Dislocation (2374)
Date of Event 03/31/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001822565-2022-01235.Product device code: phx.Concomitant medical products: item# 110031405; lot# 65172689.Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported a patient underwent a revision surgery approximately six (6) months post-implantation due to dislocation caused by disassociation of the poly bearing and tray.Attempts have been made and no further information has been provided.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
(b)(4).D10 - medical product: catalog #: 110030777, cr 40mm glenosphere +3mm cocr, lot # 65054408.Reported event was confirmed as visual examination of the returned product identified a reverse shoulder mini humeral tray and reverse shoulder bearing were returned.As returned, damage is seen articulating surface, backside, and rim feature of the bearing.The tray exhibits damage / burnishing on the inner surface and rim features.No product was returned or pictures provided for the glenosphere; visual and dimensional evaluations could not be performed.Device history record was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BEARING +3 MM THICKNESS 40 MM DIAMETER
Common Device Name
SHOULDER PROSTHESIS
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key14218526
Report Number0001822565-2022-01026
Device Sequence Number4124555
Product Code PHX
UDI-Device Identifier00887868231599
UDI-Public(01)00887868231599(17)260221(10)64862430
Combination Product (Y/N)N
Initial Reporter StateID
Initial Reporter CountryUS
PMA/510(K) Number
K181611
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/21/2026
Device Catalogue Number110031428
Device Lot Number64862430
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/07/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/31/2022
Supplement Date Received by Manufacturer06/29/2022
Initial Report FDA Received Date04/27/2022
Supplement Report FDA Received Date06/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
NI
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE NARRATIVE IN H10
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age54 YR
Patient SexMale
-
-