• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION CRE WIREGUIDED; DILATOR, ESOPHAGEAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION CRE WIREGUIDED; DILATOR, ESOPHAGEAL Back to Search Results
Model Number M00558480
Medical Device Problem Codes Leak/Splash (1354); Use of Device Problem (1670)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/02/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Facility name: (b)(6) university.(b)(4).The device has not been received for analysis.Should the device become available for analysis and there is any further relevant information, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that a cre wireguided dilatation balloon was to be used during an esophageal dilation procedure performed in the esophagus on (b)(6) 2022.During preparation, it was noted that the balloon was leaking liquid.The procedure was completed with another cre wireguided dilatation balloon.Note: the instructions for use (ifu) indicate that this balloon should not be pre-inflated prior to use in the procedure.However, the customer reported that the balloon was pre-inflated outside of the patient.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
 
Additional Manufacturer Narrative
Block e1: facility name: (b)(6).Block h6: device code a0504 captures the reportable event of balloon leaked in the esophagus.Block h10: investigation result the returned cre wireguided dilatation balloon was analyzed and a visual examination found the catheter of the device has no damages.Visual and microscopic examination of the balloon found it torn longitudinally.With all the available information, boston scientific concludes the reported event was confirmed.The customer pre-inflated the balloon before inserting the device into the working channel and this manipulation can provoke damages to the balloon.It is possible that interaction with any surface during the procedure could create friction on the balloon, causing the problem of torn longitudinally.Therefore, the most probable root cause is adverse event related to procedure.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.A labeling review was performed and from the information available, this device was not used per the instructions for use (ifu)/product label.It was stated in the instructions for use (ifu) that this balloon should not be pre-inflated prior to use in the procedure.However, the customer reported that the balloon was pre-inflated outside of the patient.
 
Event or Problem Description
It was reported to boston scientific corporation that a cre wireguided dilatation balloon was to be used during an esophageal dilation procedure performed in the esophagus on april 2, 2022.During preparation, it was noted that the balloon was leaking liquid.The procedure was completed with another cre wireguided dilatation balloon.Note: the instructions for use (ifu) indicate that this balloon should not be pre-inflated prior to use in the procedure.However, the customer reported that the balloon was pre-inflated outside of the patient.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRE WIREGUIDED
Common Device Name
DILATOR, ESOPHAGEAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
cork business technology park
model farm road
cork
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key14218878
Report Number3005099803-2022-02211
Device Sequence Number4208265
Product Code KNQ
UDI-Device Identifier08714729339410
UDI-Public08714729339410
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K110833
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/16/2024
Device Model NumberM00558480
Device Catalogue Number5848
Device Lot Number0027324353
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/02/2022
Supplement Date Received by Manufacturer07/13/2022
Initial Report FDA Received Date04/27/2022
Supplement Report FDA Received Date08/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/17/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age73 YR
Patient SexFemale
Patient Weight68 KG
-
-