• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US 23MM HEX DRIVER W/TEARDROP HANDLE -NS; SURGICAL SCREWDRIVER, REUSABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US 23MM HEX DRIVER W/TEARDROP HANDLE -NS; SURGICAL SCREWDRIVER, REUSABLE Back to Search Results
Model Number 270127
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi:(b)(4).The lot number was unknown.The lot number was unknown; therefore, the expiration date, manufacturing site name and device manufacture date were unknown.
 
Event or Problem Description
It was reported by the sales rep that during an unknown procedure on (b)(6) 2022, it was observed that the tip of the 23 mm hex driver with teardrop handle was bent and not allowing the screw to fit on the tip.Another like device used to complete the procedure.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary: the product was returned to depuy synthes mitek for evaluation.The depuy synthes mitek team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that 23mm hex driver w/teardrop handle -ns was bent on the distal tip.A dimensional inspection was unable to be performed due to post-manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended and/or excessive forces.As part of depuy synthes mitek quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the 23mm hex driver w/teardrop handle -ns would contribute to the complained device issue.During the investigation, no product design or manufacturing issues were observed that may have contributed to the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent to which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
23MM HEX DRIVER W/TEARDROP HANDLE -NS
Common Device Name
SURGICAL SCREWDRIVER, REUSABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key14218907
Report Number1221934-2022-01238
Device Sequence Number17178077
Product Code LXH
UDI-Device Identifier10886705014033
UDI-Public10886705014033
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number270127
Device Catalogue Number270127
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/11/2022
Supplement Date Received by Manufacturer05/06/2022
05/10/2022
Initial Report FDA Received Date04/27/2022
Supplement Report FDA Received Date05/09/2022
05/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-