• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. TRULIANT PS CEM FEM PS CEM RIGHT SZ 6; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. TRULIANT PS CEM FEM PS CEM RIGHT SZ 6; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED Back to Search Results
Model Number TRULIANT PS CEM FEM PS CEM RIGHT SZ 6
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 04/07/2022
Type of Reportable Event Serious Injury
Event or Problem Description
As reported, approximately one year post initial right tka, this (b)(6) male patient had revision due to loose femur and complained of pain.Patient was revised with competitor's implants.There was no breakage of device or surgical delay/prolongation.Patient was last known to be in stable condition following the event.Sales rep was unable to obtain photos/x-rays.The device will be return.
 
Additional Manufacturer Narrative
Concomitant medical products truliant tib imp ps insert sz 6 9mm (cat# 02-022-35-6009 / serial(b)(4)).Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Additional Manufacturer Narrative
(h3) the revision reported was likely the result of an insufficient bond between the femoral component and the bone, which led to aseptic (non-infected) femoral loosening and pain.The most probable root cause associated with the reported event of "loosening - femoral¿ is associated with weakened integration of the femoral component at the bone-implant interface due to loss of fibrous and/or bony tissue and leading to compromised anchorage of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRULIANT PS CEM FEM PS CEM RIGHT SZ 6
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
kate jacobson
2320 nw 66 court
gainesville, FL 32653
3523771140
MDR Report Key14219059
Report Number1038671-2022-00475
Device Sequence Number5389348
Product Code JWH
UDI-Device Identifier10885862314598
UDI-Public10885862314598
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K170240
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTRULIANT PS CEM FEM PS CEM RIGHT SZ 6
Device Catalogue Number02-020-11-0360
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/20/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/07/2022
Supplement Date Received by Manufacturer10/03/2022
Initial Report FDA Received Date04/27/2022
Supplement Report FDA Received Date10/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/04/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
Patient SexMale
-
-