• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDICAL COMPONENTS, INC. 2.6F X 50CM DUAL VASCU-PICC W/MST

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDICAL COMPONENTS, INC. 2.6F X 50CM DUAL VASCU-PICC W/MST Back to Search Results
Model Number MD170126024S
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/04/2022
Type of Reportable Event Malfunction
Event or Problem Description
Dressing was wet (no indication in chart- just said possible displacement in the event note) then an order to replace the line.The hole was just above the hub.
 
Additional Manufacturer Narrative
An investigation has been initiated.Additional information has been requested from the facility.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
The device involved in the incident was not returned for evaluation.A photograph provided shows what appears to be a ruptured lumen between the hub and 1st lumen marking.The contract manufacturer conducted a review of the manufacture records for the lot number reported.Their investigation revealed the device was manufactured and inspected according to specification with no non-conformances or abnormalities.The manufacture process includes a 100% leak test performed as a last step in the process.This test would have detected any leaks, holes, or weak spots if it had existed at the time of manufacture.Without an evaluation of the device a root cause cannot be determined.The instructions for use (ifu) contains the following cautions and warnings: 1.Do not use infusion equipment which can exceed the working pressure of 1.0bar max/750mmhg (14.5 psi).2.Only use infusion equipment complying with standards, which do not exceed a shut-off pressure of 1.0bar.3.Bolus injections should be slow and must not exceed the maximum bolus pressure of 1.2bar/900mmhg (17.4 psi).4.Do not use high-pressure injectors for contrast medium studies.Excessive pressures may damage catheter.5.Small syringes will generate excessive pressure and may damage the catheter.The use of 10cc or larger syringes are recommended.6.Do not use sharp instruments near the extension lines or catheter lumen.7.Do not use scissors to remove dressing.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.6F X 50CM DUAL VASCU-PICC W/MST
Common Device Name
VASCU-PICC
Manufacturer (Section D)
MEDICAL COMPONENTS, INC.
1499 delp drive
harleysville PA 19438
Manufacturer (Section G)
MEDICAL COMPONENTS, INC.
1499 delp drive
harleysville PA 19438
Manufacturer Contact
lynn winkler
1499 delp drive
harleysville, PA 19438
2152564201
MDR Report Key14219144
Report Number2518902-2022-00029
Device Sequence Number4124594
Product Code LJS
UDI-Device Identifier00884908147230
UDI-Public884908147230
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date04/30/2024
Device Model NumberMD170126024S
Device Catalogue NumberMD170126024S
Device Lot NumberMQBD360
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/26/2022
Supplement Date Received by Manufacturer04/26/2022
Initial Report FDA Received Date04/27/2022
Supplement Report FDA Received Date05/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/28/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-