| Model Number |
928853 |
| Medical Device Problem Code |
Missing Information (4053)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/04/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported 2 bd¿ insulin syringes were missing the expiration date on the packaging.The following information was provided by the initial reporter: "consumer reported not seeing the exp date on this box and bags.
|
| |
|
Additional Manufacturer Narrative
|
|
Medical device expiration date: na.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
The following fields were updated due to additional information: d9: device available for eval yes.D9: returned to manufacturer on: 22-jun-2022.Investigation summary: customer returned a single syringe in a polybag labeled for 1ml, 30 gauge, 8mm syringe from lot 0168471.The syringe was inspected and measured to ensure that using it was as least harmful as intended.The outer diameter the syringe needle was measured to be 0.0124 in, which is within acceptable outer diameters for 30 gauge needles (0.0120 in to 0.0125 in).An unopened polybag for 1ml, 30 gauge, 8mm syringes from lot 7331982 was also returned.No defects were observed with these syringes and associated polybag.Additionally, no shelf carton was returned for inspection with these samples.No syringes, polybags, or shelf cartons from lot 8071585 were returned.A review of the device history record was completed for batch# 8071585.All inspections were performed per the applicable operations qc specifications.There were zero (0) notifications noted that pertained to the complaint.Expiration dates were not printed on packaging until 08nov2019.Embecta was unable to confirm the customer¿s indicated failure of label info missing as the samples received were not from the lot in question.Per manufacturing, dates not printed on polybags until nov, 2019.Root cause cannot be determined at this time as the issue is unconfirmed.H3 other text : see h10.
|
| |
|
Event or Problem Description
|
|
It was reported 2 bd¿ insulin syringes were missing the expiration date on the packaging.The following information was provided by the initial reporter: "consumer reported not seeing the exp date on this box and bags.
|
| |
|