The reported device was returned, and the reported event was confirmed.Based on the evaluation, the device malfunction did occur.However, there is no existing nonconformance/capa/hhe/d/ie/product holds against the reported product that did or could cause or contribute to the reported event.Monthly post market trending review identified no adverse trends or actionable trends for the reported event or device.Zimvie quality management system (qms) has controls in place to ensure the distribution of conforming product.Therefore, based on the available information, the product was within specifications and conforming when it left zimvie.Dhr review was completed for the subject lot number.It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Complaint history review was performed for the reported lot number for similar events and one (1) other complaint was identified.Functional testing to recreate the reported event; verified the implant body could not be removed from the fixture mount.The complaint is related to the functional performance of the device.A definitive root cause could not be identified.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimvie will continue to monitor for trends.
|